  ZG116 SERIES - ZG116 / ZG116A-M  
   
 
ZG116 / ZG116A-M
ZG116 / ZG116A-M
ZG1163 / ZG1163R (RH)
 
 
  1. Enter Button
  6. LCD Display
  2 .Down Button
  7 .Gas Entry Hole
  3. Up Button
  8. Power inlet
  4. Mode Button
 

9. RJ45 socket
(only for factory use)

  5. Alarm LED
 
 
 
 
Features & Display Features and Modes
 
 
Ultra low cost CO2 Monitor achieved by New IR-SoC Technology
It uses NDIR technology to improve the long term stability
Portable, accurate handheld CO2 Monitor
The Alarm mode will sound at 1000 PPM or adjust the alarm level
Max / Min mode
 
 
Mode Funtion
 
 
  There are several Modes which we can adjust the setting parameters. These modes are Alti Mode,Alarm Mode, Outside Mode, Ucal Mode, Datalogger Mode, MaxMin Mode and rcFS Mode in sequence.
 
 
  Alarm icon, the sound alarm can be turned on/off by user.
ALTI
  Compensate the pressure changes with appropriate altitude of location when measure.
ALARM
  Alarm while CO2 concentration exceed one threshold, Alarm sound can be on or off.
OUTSIDE
  Modify the outside CO2 concentration,for defining the ventilation rate.
CALI
  Calibrate the sensor while the reading deviates from the actual CO2 concentration.
DATALOGGER
  Show the past CO2 and Temperature records in the past 24 hours.
MaxMin
  Show the Max and Min CO2 reading before being cleared or powered off after Power On.
rcFS
  Recover the factory setting, if the device cannot show accurate reading.
 
 
Safety Installation Step
 
 
Warning: Your safety is very important to us .To ensure to use the product correctly and safely, we would like to
    draw your attention to read the warning and entire User Manual before using the product. These are
    important safety information and should be observed at all times.
1.
Please take the devices lightly, Do not subject the product to impact or shock.
2.
Do not immerse the product in water. Water can cause electric shock, fire or malfunction which may result in
  damage.
3.
Do not submerge the product into water or any liquid. Water may cause electric shock, fire or malfunction
  which may result in damage.
4.
Only use the included power adaptor. Improper power adaptor or power sources may cause serious damage
  to the product, or result in injury or death to the user.
 
 
Caring for your product
 
 
To ensure you receive the maximum benefit from using this product, please observe the follow guidelines.
1.
Cleaning----Disconnect the power before clean. Use a damp cloth, do not use the liquid cleaning agent, such
                        as benzene, thinner or aerosols.    
2.
Repair----Do not attempt to repair the product or modify the circuitry by yourself. Please contact with our local
                  dealer or a qualified repairman if the product needs servicing.
3.
Calibration--- The vents allow the air circulation liquid for measurement of the CO2 concentration and the
                         ventilation and should not be blocked.
 
 
Customizing Settings
 
 
  There are several Modes which we can adjust the setting parameters. These modes are Alti Mode,Alarm Mode, Outside Mode, Ucal Mode, Datalogger Mode, MaxMin Mode and rcFS Mode in sequence.

Using the OUTSIDE mode

 
1.Press Mode, the OUTSIDE flashes.
2.Press Enter, shows OUTSIDE, CO2 and ppm flashes. Press the Up/Down to adjust the data.
3.Press Enter to save the setting. After the modification, the VENT Rate will change.
*Note: Ventilation rate represents how much air is introduced into the indoor space from outside. Low numbers
  indicate low ventilation rates and potentially, poor air quality. High levels indicate excessive ventilation and potential excessive energy usage. To obtain an accurate measurement, reading should be taken 2~3 hours after occupancy has stabilized in a space or at a peak in daily CO2 concentrations. For indoor air quality control, CO2 value is an indicator of ventilation rate. 400ppm (Parts Per Million) is the default CO2 concentration outside (according to ASHRAE: American Society of Heating, Refrigeration and Air conditioning Engineers.

Using the CALI(calibration) mode

 
1.Press Mode, The CALI flashes.
2.Press Enter, the CALI icon on the display. Adjust the CO2 reading by Up/Down button.
3.Press Mode for more than 10sec,CALIBRATING flashes. Calibration will be done automatically after 10
   minutes and the LCD will display "Pass" or "Fail". If it shows "Fail", please try again.

Using the DATALOGGER mod

 
1.Press Mode, DATALOGGER flashes.
2.Press Enter, CO2 reading and Temperature show on the LCD.
3.Press Up/Down to page up/down the reading.

*Note: With the Built-in Datalogge,ZG116 can provide the past CO2 and temperature reading within the past 24

            hours.

Using the MAX MIN mod

 

 
1.Press Mode, the MAX MIN flashes.
2.Press Enter, the MAX and MIN CO2 reading alternatively shows.
3.Press Down and the "CLr" will flash on the LCD. Press Enter to CLEAR the MAX and MIN record.

*Note: With the Built-in MAX MIN mode, ZG116 can provide the Maximum and Minimum CO2 readings since the

            device has last been turned on.

Using the "rcFS" Mode

 
 
1.Press Mode, The rcFS flashes.

2.Press Enter, the icon “no” shows on the LCD, Press the Up/Down to select the no/yes.
3.After selecting, press Enter to save the setting.

*Note: If the user sets the data or calibrates the ZG116 incorrectly, use the rcFS (recover the factory Setting) to

            come back the default factory setting.
 
 
Specifications
 
 
  Method - NDIR
  Display - LCD
  Independent CO2 and Temperature readings.
  Calculates and Displays Ventilation Rates
 

Sample Method - Diffusion or flow through (50 ~200 ml/min)

 
 
Performance - CO2
  ZG116 ZG116A-M
Measurement Range 0~3,000ppm 0~10,000ppm
Resolution
1ppm at 0~1,000ppm;
5ppm at 1,001~2,000ppm;
10ppm at 2,001~3,000ppm
1ppm at 0~1,000ppm;
5ppm at 1,001~2,000ppm;
10ppm at 2,001~3,000ppm
20ppm at 4,001~6,000ppm;
40ppm at 6,001~10,000ppm;
Accuracy 0~2,000ppm: ±70 ppm or ±5% of reading, whichever is greater ;  over 2000ppm: +/-7%
Repeatability  ±20 ppm
Temperature Dependence ZG116-F:
±0.2% of reading per °C or ±2 ppm per °C, whichever is greater, refer­enced to 15°C-35 °C. Otherwise ±0.4 % of reading per °C or ±4 ppm per °C at other temperature range, whichever is greater, referenced to 25°C  

Others:
±0.1% of reading per °C or ±2 ppm per °C, whichever is greater,  refer­enced to 25°C
Pressure Dependence 0.13% of reading per mm Hg
(Corrected via user input for altitude)
Response Time <2min for 63% of step change
Warm-Up Time <60 seconds at 22°C
Performance - Temperature
  Temperature Range: Display 32 to 122°F (0 to 50°C)
  Display Resolution: 0.1°F (0.1°C)
  Display Options: °F / °C, or Off.  Set with Up / Down button
  Accuracy: ±1°C( ±2°F) When the fan blows to the device directly, the accuracy of
  temperature is + / -1.5 degC
  Response Time: 20-30 minutes (case must equilibrate with environment)
General Operating Conditions
  Operating Temperature: 0°C to 50°C (32°F to 122°F); 0 ~ 95% RH non-condensing
  Storage Temperature: -4 to 140°F (-20 to 60°C)
Power Supply
  6 VDC from external AC/DC adapter which is included in package (Use specified AC
  adapter only)
  DC Output voltage : 0.1°F (0.1°C)
  AC Input voltage & Frequency: 100 ~ 240 VAC   50 / 60 Hz
Dimensions
  123.5 x 88 x 46.05 mm
Weight (including batteries):
  162.3 grams
LCD ERROR MESSAGES
  This section includes a list of Frequently Asked Questions for problems you may encounter with the ZG116 CO2 Monitor.
Fault Icon
Description of
the fault
Suggested Actions
  The ambient temperature has
  exceeded the operating temperature range 0°C
  to 50°C(32°F to 122°F)
  This error will clear when the
  temperature returns to the range
  between 0°C to 50°C (32°F to 122°F)
  EEPROM System Problem
  Please reconnect AC adapter to ZG116 CO2 Monitor. If the "Err5" still
  appears, please contact the Service
  Department for further assistance.
Kinds of Adapters for your option
ACAeu
(for European ,
including
GermanItaly )
ACAgb
(for British,
including
Singapore、
BS1363 )
ACAau
(for Australia
New Zealand)
ACAus
(for American ,
including
JP、CN)
Available Packages
  Dimensions: 123.5*88*46.05mm
  Weight :162.3g
  Power Adaptor Weight :
  U.K Spec. :67.3g
  EU Spec. : 60.6g
  U.S Spec. : 56.7g
  AU Spec. : 93.2g
   
  One Set Included:
 
1 CO2 Monitor
 
1 Power Adaptor
 
1 User Manual
 
1 White Box
 
1 Anti-ESD PE Bag (Pink)
 
1 PE Bag ( Package Power
    Adaptor)
  One Carton Included:
 
Q'ty: 40pcs PaperBoxes
 
White Box Size: 146*122*103mm
 
Carton Size:
    645(L)*312(W)*436(H)mm;
    3.09 Cube ft
     
 
N.W:
 
U.K Spec. Power Adaptor: 14.2 kgs
 
EU Spec. Power Adaptor: 14.0 kgs
 
U.S Spec. Power Adaptor: 13.8 kgs
 
AU Spec. Power Adaptor: 15.4 kgs
 
G.W:
 
U.K Spec. Power Adaptor: 16.7 kgs
 
EU Spec. Power Adaptor: 16.5 kgs
 
U.S Spec. Power Adaptor: 16.3 kgs
 
AU Spec. Power Adaptor: 17.9 kgs
 
 

TOTAL: