ZGw08 series: Wall mount CO2 monitor
- use in Indoors or relay to HVAC system for Energy Saving

 
 
ZGw0806 SERIES (Under Development )
ZGw08VRC
Red LED display:
shows the current Indoor Air Quality situation
LED >1000 ppm
m
 Buzzer Alarm : 70dB@10 cm
Available Packages
Specifications
Features & Display Features and Modes
LCD Display Symbol

Safety Installation Step

Caring for your product
Installation Step
Dimension
Customizing Settings
Calibration & Warranty

View
 A. Main LCD Display  F. Enter Button
 B. Red LED Display   (>1000ppm)  G. Screw Position
 C. Mode Button  H. Power Inlet
 D. Up Button  I. RJ45 Socket
 E. Down Button  J. Gas Entry Hole

Available Packages:
One Outter Carton Included:
Inner Carton Size:
1 set Inner Carton:
Total Q'ty:
Outter Carton Size:
Cube ft:
N.W:
G.W:
 
Display Features and Modes
ZyAura_CO2_Monitor_Carbon_Dioxide_ZG106monitor
Features:
NDIR ( Non-Dispersion-Infrared) technology used to measure
  CO2 concentration.
Red LED display shows the current Indoor Air Quality situation.

Reliable Sensor provides long-term calibration stability .
Visual and audible alarm function can be adjusted by the user.

LCD Display Symbol

Symbol
Meaning
Description
LCD01
CO2 Concentration PPM
Parts Per Million
The current CO2 concentration in your household.
LCD02
Ventilation Rate cfm/p
Cubic Feet Minute/Person
The Present Cubic Feet/Minute
Ventilation for one person.
LCD03
Ventilation Rate L/P/S
Liter/Person/Second
The Current Liter/ Second
Ventilation Rate for one person.
Temperature(Celsius) The present temperature in your household.
Temperature (Fahrenheit) The current temperature in your household.
LCD04
Alarm

The Alarm Setting Icon.

Altitude Compensate the pressure changes with appropriate altitude of location when measure
Alarm The alarm level is adjustable
Outside Modify the outside CO2 concentration for defining the ventilation rate
Calibration To calibrate the CO2 sensor when the accuracy deviates from the actual CO2 concentration
Dataloggor Display the past CO2 and Temperature records in the past 24 hours
Max Min Display the Max and Min CO2 reading record before being cleared or the device is powered off
Clear To clear the Max and Min CO2 reading record
Recover Factory Setting To Recover Factory Setting to cancel the customizing Setting

Safety Installation Step

Warning: Your safety is very important to us. To ensure that you use your product correctly and safely, we would like to draw your attention to read the warning labels and the User Manual before using the product. These are important safety information and should be observed at all times.
1. Please handle the device lightly; do not subject the product to impact or shock.
2. Do not immerse the product in water.
3. Do not touch the exposed electronic circuitry of the device under any circumstances.
  Keep the circuit closed during installation, as there is the danger of electric shock.
4. Please use only the included power adaptor. Improper power adaptors or power sources can cause   serious damage to the product, or result in injury or death to the user.
5. Do not keep the product under hot and/or extremely moist environment.
  Keep the product away from any heat source or near water.

Caring for your product

To ensure you receive maximum benefit from using this product, please observe the follow  guidelines.
1. Cleaning--Disconnect the power before cleaning. Use a damp cloth; do not use liquid cleaning  agents, such as benzene, thinner or aerosols.    
2. Repair--Do not attempt to repair the product or modify the circuitry by yourself. Please contact our  local dealer or a qualified repairman if the product needs servicing.
3. Calibration-- Please observe the calibration operation to make sure the accuracy for the devices  when necessary.
4. Air circulation--The vents allow the air circulation liquid for measurement of the CO2 concentration  and the ventilation and should not be blocked.

Installation Step
Before installation, please carefully remove all the parts from the box.
Step 1: Release the screw from the device by screwdriver, and take the front cover off.
        
Step 2: Release the four screws from the back cover, take the CO2 board from the back cover.
        
Step 3: Separate the Front Cover from the CO2 Board and the Back Cover.
    
Step 4: Use the screwdriver to fix the back cover
     to the wall.
    
Step 5: Re-assemble the CO2 board by     screwdriver to the back cover.
    
Step 6: Install the front cover to the CO2 board     after the CO2 board is fixed well by screws.
    
Step 7: Tighten the Front Cover screw with     screwdriver.
    

Customizing Settings

When the power has been connected, The ZGw0806 CO2 monitor will begin to work. In order to meet your personal requirements, it is advisable to set up the customizing parameters.

Setting the alarm function

setting01
1.Press MODE, the speaker icon flashes simultaneously.
2.Press ENTER, use up/down to select the on/off.
3.Press ENTER again to save the data.

Setting the ALTI mode

setting01
1.Press MODE ,the ALTI icon flashes.
2.Press ENTER, ALTI show on the display And press the mode button to alter between the m(meter)  and ft(feet).
3.Adjust altitude (step=100m/500ft) by up/down button.
4.Press ENTER again to save the data.

Setting the Alarm parameter Mod

setting01
1.Press MODE, the speaker icon flashes.
2.Press ENTER. Using the Up/Down to set the parameter.
3.Press ENTER again to save the data.

Note: When CO2 Alarm level is ≧1,000 ppm, the interval is ±100ppm, when the CO2 Alarm level is <1,000 ppm, the interval is ±50 ppm.

Using the OUTSIDE Mode

setting01
1.Press MODE, The OUTSIDE flashes.
2.Press ENTER, shows OUTSIDE,CO2 and ppm flashes. Press the Up/Down to adjust the reading.
3.Press ENTER to save the setting. After the modification, the VENT Rate will change.

Note: Ventilation rate represents how much air is introduced into the indoor space from outside. Low numbers indicate low ventilation rates and potentially, poor air quality. High levels indicate excessive ventilation and potential excessive energy usage. To obtain an accurate measurement, reading should be taken 2~3 hours after occupancy has stabilized in a space or at a peak in daily CO2 concentrations. For indoor air quality control, CO2 value is an indicator of ventilation rate. 400ppm (Parts Per Million) is the default CO2 concentration outside (according to ASHRAE: American Society of Heating, Refrigeration and Air conditioning Engineers).

Using the CALI (calibration) Mode

setting01
1.Press MODE, The CALI flashes.
2.Press ENTER, the CALI icon on the display. Adjust the lower display to the ambient CO2 reading by  Up/Down button.
3.Press MODE for more than 10sec, CALIBRATION flashes. Calibration will be done automatically after  5 minutes and the LCD will display “Pass” or “Fail.” If it shows “Fail,” please try again.
 

Using the Datalogger Mode

setting01
1.Press MODE, DATALOGGER flashes.
2.Press ENTER, the reading of CO2 and Temperature shows on the LCD.
3.Press Up/Down to page up/down the reading.

Remark: With the Built-in Datalogge,ZGw0806 can provide the past CO2 and temperature reading within the past 24 hours.

Using the MAX MIN Mode

setting01
1.Press MODE, the MAX MIN flashes simultaneously.
2.Press ENTER, the reading of CO2 and Temperature shows on the LCD alternatively.
3.Press Up/Down and the “CLR” will flash on the LCD. Press Enter to CLEAR the MAX and MIN record.

Note: With the Built-in MAX MIN mode, ZGw0806 can provide the Maximum and Minimum CO2 reading record after the device is powered on. If users press up/down to clear the MAX and MIN CO2 reading record, ZGw0806 will provide the new MAX and MIN CO2 reading from that time.

Using the "rcFS" Mode

setting01
1.Press MODE, The rcFS flashes simultaneously.
2.Press ENTER, the icon “no” shows on the LCD, Press the Up/Down to select the no/yes.
3.After selecting, press ENTER to save the changes.

Note :If the users set the data or calibrate the sensor wrongly. You can use the rcFS(Recover the factory Setting) to come back to factory setting data.

 
 
 

SPECIFICATIONS
CO2 Specification
Item
ZGw0806
ZGw0806-M
ZGw0806-H
Measurement Range
0 ~ 3,000ppm display
0 ~ 10,000ppm display
0 ~ 50,000ppm display
Display Resolution 1ppm
Accuracy ±75ppm or ±5% of reading whichever is greater
Repeatability ±20 ppm @400ppm
Annual Drift <20ppm/Year
Temperature Dependence ±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, refer­enced to 25°C
Pressure Dependence 0.13% of reading per mm Hg (Corrected via user input for altitude)
Response Time <2min for 63% of step change
Warm-Up Time <60 seconds at 22°C
Sound Alarm 70db@10cm
Zone LED Display Red: >1000ppm
Power Supply 6VDC AC adapter
Storage Temperature -4°F to 158°F (-20°C to +70°C)
Temperature specification
Temperature Range Display 32°F to 122°F (0°C to 50°C)
Display Resolution ±20 ppm @400ppm
Display Options °F / °C
Accuracy ±2°F (±1°C)
Response Time 20-30 minutes (case must equalize with environment)
Operating Conditions
Operating Temperature 32°F to 122°F (0°C to 50°C )
Humidity Range 0 ~ 95% RH non-condensing

Dimension

Calibration & Warranty

Note: Before calibrating, you need to obtain standard gas or semi-standard gas, there are 3 ways to get standard gas

Method A: use CO2 in office/building

-Use two meters (One is the device for calibration. The other one is a calibrated (new) one.

-Use ambient room gas for calibration in office, waiting at least 10min, until the CO2 reading doesn’t change.
(Notice: user must not breathe toward the ZGw0806, CO2 from the user will affect the reading of ZGw0806)

-Use the reading of the new device as the standard reading.
-Calibrate device by the Cali Mode instruction.

Method B: use CO2 outsides

-Use ambient room gas for calibration outside. Wait at least 10min, until the CO2 reading stabilizes.
(Note): user must not breathe toward the ZGw0806, CO2 from the user will affect the reading of ZGw0806)

-Use 380~420ppm as the standard reading.
-Calibrate the device by the Cali Mode instruction.

Method C: use standard CO2 in the bottle

- Pump the standard CO2 gas(0~1000ppm, flux = 0.1~0.2 liter/min) into the ZGw0806 from the Gas Entry Hole (top left corner ) waiting about 2~3min.
- Calibrate the device by the Cali Mode instruction.

 

 
     
opyright (c) 2006-2007 RADIANT INNOVATION INC. All Rights Reserved.

English: (in order of frequency; CO2 monitor is the best one)CO2 Monitors & Analyzers ,Gas analyzer ,carbon dioxide ,CO2,CO2 analyze, Gas monitor,co2 moniotr, Carbon Dioxide monitor,concentration,ppm,dangerous,ventilate,ventilation rate, air supply ratestemperature,monitor,air quality,portable monitor,Testing Equipment,measuregas detection, CO2 Meter,indoor air quality,IAQ, 380ppm ,400ppmASHRAE, HVAC,OSHADemand Controlled Ventilation,DCVFrench: Moniteur de CO2 , Anhydride carbonique, Analyseur de gaz, qualité d'intérieur d'air,German: CO2-Monitor, Kohlendioxyd, Konzentration, Innenluftqualität, Maß,Greek: Όργανο ελέγχου του CO2, Διοξείδιο του άνθρακα, μέτρηση, εσωτερική ατμοσφαιρική ποιότηταHugarian: CO2 idegen rádióadást figyel, Gáz Elemző, szellőzés, légcsere, összpontosítás, koncentrációIceland:  CO2 skjár, Gas Greina í þætti,samsöfnun, samdráttur, einbeitingItalian: Video del CO2, Anidride carbonica, Concentrazione, ventilazione, qualità dell'interno dell'ariaNetherlands (Nederland):  De monitor van Co2 , De analysator van het gas , Kooldioxide ,ConcentratieNorway: CO2 dataskjerm, Gass Analyserer, konsentrasjonen, ventilasjonen, Indoor Luften Kvalitet, konsentrasjon,Finish: CO2 järjestäjä , Bensiini Analysoiminen , Hiili Dioksidi ,ilmanvaihto, tuuletusPolish: wentylacja, przewietrzenie, Żyrować Przewietrzyć Różnice jakościowe, Dwutlenek węglaPortuguese: Monitor do CO2, Dióxido de carbono, Concentração, ventilaçãoRussian: Монитор CO2, Двуокись углерода, Вентиляция,Концентрация, Indoor качество воздухаSpanish: Minitor del CO2, Bióxido de carbono, Ventilación, ConcentraciónSwedish: CO2 övervakar , Gasa övervakar ,Koldioxid , Lufta inomhus kvalitets-Japanese: 二酸化炭素のモニター,ガス分析器,換気,濃度,屋内空気質Korea: 이산화탄소 감시자, 가스 해석기, 농도, 실내 공기 질Chinese(Simpified): CO2 检测仪,二氧化碳检测仪,通风率,浓度Chinese(Traditional): CO2 檢測儀,二氧化碳檢測儀,通風率,濃度